Articles producció científica> Ciències Mèdiques Bàsiques

Calibration and performance of a real-time gamma-ray spectrometry water monitor using a LaBr3(Ce) detector

  • Identification data

    Identifier: imarina:5132681
    Handle: http://hdl.handle.net/20.500.11797/imarina5132681
  • Authors:

    Prieto E., Casanovas R., Salvadó M.
  • Others:

    Author, as appears in the article.: Prieto E., Casanovas R., Salvadó M.
    Department: Ciències Mèdiques Bàsiques
    URV's Author/s: Casanovas Alegre, Ramon / Salvadó Artells, Marçal
    Keywords: Scintillation gamma-ray spectrometry Monte carlo simulation Mdac Labr (ce) 3 Efficiency calculation
    Abstract: A scintillation gamma-ray spectrometry water monitor with a 2″ × 2″ LaBr3(Ce) detector was characterized in this study. This monitor measures gamma-ray spectra of river water. Energy and resolution calibrations were performed experimentally, whereas the detector efficiency was determined using Monte Carlo simulations with EGS5 code system. Values of the minimum detectable activity concentrations for 131I and 137Cs were calculated for different integration times. As an example of the monitor performance after calibration, a radiological increment during a rainfall episode was studied.
    Thematic Areas: Zootecnia / recursos pesqueiros Saúde coletiva Radiation Química Physics, atomic, molecular & chemical Odontología Nutrição Nuclear science & technology Medicina veterinaria Medicina ii Medicina i Materiais Interdisciplinar Geografía Geociências Farmacia Ensino Engenharias iv Engenharias iii Engenharias ii Engenharias i Ciências biológicas iii Ciências biológicas ii Ciências biológicas i Ciências ambientais Ciências agrárias i Ciência de alimentos Chemistry, physical Biotecnología Biodiversidade Astronomia / física Antropologia / arqueologia
    licence for use: https://creativecommons.org/licenses/by/3.0/es/
    ISSN: 0969806X
    Author's mail: ramon.casanovas@urv.cat m.salvado@urv.cat
    Author identifier: 0000-0002-3954-158X 0000-0002-9399-6649
    Record's date: 2023-02-18
    Papper version: info:eu-repo/semantics/acceptedVersion
    Link to the original source: https://www.sciencedirect.com/science/article/abs/pii/S0969806X17306369?via%3Dihub
    Licence document URL: http://repositori.urv.cat/ca/proteccio-de-dades/
    Papper original source: Radiation Physics And Chemistry. 144 444-450
    APA: Prieto E., Casanovas R., Salvadó M. (2018). Calibration and performance of a real-time gamma-ray spectrometry water monitor using a LaBr3(Ce) detector. Radiation Physics And Chemistry, 144(), 444-450. DOI: 10.1016/j.radphyschem.2017.10.008
    Article's DOI: 10.1016/j.radphyschem.2017.10.008
    Entity: Universitat Rovira i Virgili
    Journal publication year: 2018
    Publication Type: Journal Publications
  • Keywords:

    Chemistry, Physical,Nuclear Science & Technology,Physics, Atomic, Molecular & Chemical,Radiation
    Scintillation gamma-ray spectrometry
    Monte carlo simulation
    Mdac
    Labr (ce) 3
    Efficiency calculation
    Zootecnia / recursos pesqueiros
    Saúde coletiva
    Radiation
    Química
    Physics, atomic, molecular & chemical
    Odontología
    Nutrição
    Nuclear science & technology
    Medicina veterinaria
    Medicina ii
    Medicina i
    Materiais
    Interdisciplinar
    Geografía
    Geociências
    Farmacia
    Ensino
    Engenharias iv
    Engenharias iii
    Engenharias ii
    Engenharias i
    Ciências biológicas iii
    Ciências biológicas ii
    Ciências biológicas i
    Ciências ambientais
    Ciências agrárias i
    Ciência de alimentos
    Chemistry, physical
    Biotecnología
    Biodiversidade
    Astronomia / física
    Antropologia / arqueologia
  • Documents:

  • Cerca a google

    Search to google scholar