TÍTULO:
Multi-institutional investigation into the robustness of intra-cranial multi-target stereotactic radiosurgery plans to delivery errors. - imarina:9370828

Autor/es de la URV:Hernandez Masgrau, Victor
Autor según el artículo:May L; Hardcastle N; Hernandez V; Saez J; Rosenfeld A; Poder J
Direcció de correo del autor:victor.hernandez@urv.cat
Identificador del autor:0000-0003-3770-8486
Año de publicación de la revista:2024
Tipo de publicación:Journal Publications
Referencia de l'ítem segons les normes APA:May L; Hardcastle N; Hernandez V; Saez J; Rosenfeld A; Poder J (2024). Multi-institutional investigation into the robustness of intra-cranial multi-target stereotactic radiosurgery plans to delivery errors.. Medical Physics, 51(2), 910-921. DOI: 10.1002/mp.16907
Referencia al articulo segun fuente origial:Medical Physics. 51 (2): 910-921
Resumen:BackgroundThe use of modulated techniques for intra-cranial stereotactic radiosurgery (SRS) results in highly modulated fields with small apertures, which may be susceptible to uncertainties in the delivery device.PurposeThis study aimed to quantify the impact of simulated delivery errors on treatment plan dosimetry and how this is affected by treatment planning system (TPS), plan geometry, delivery technique, and plan complexity. A beam modelling error was also included as context to the dose uncertainties due to treatment delivery errors.MethodsDelivery errors were assessed for multiple-target brain SRS plans obtained through the Trans-Tasman Radiation Oncology Group (TROG) international treatment planning challenge (2018). The challenge dataset consisted of five intra-cranial targets, each with a prescription of 20 Gy. Of the final dataset of 54 plans, 51 were created using the volumetric modulated arc therapy (VMAT) technique and three used intensity modulated arc therapy (IMRT). Thirty-five plans were from the Varian Eclipse TPS, 17 from Elekta Monaco TPS, and one plan each from RayStation and Philips Pinnacle TPS. The errors introduced included: monitor unit calibration errors, multi-leaf collimator (MLC) bank offset, single MLC leaf offset, couch rotations, and collimator rotations. Dosimetric leaf gap (DLG) error was also included as a beam modelling error. Dose to targets was assessed via dose covering 98% of planning target volume (PTV) (D98%), dose covering 2% of PTV (D2%), and dose covering 99% of gross tumor volume (GTV) (D99%). Dose to organs at risk (OARs) was assessed using the volume of normal brain receiving 12 Gy (V12Gy), mean dose to normal brain, and maximum dose covering 0.03cc brainstem (D0.03cc). Plan complexity was also assessed via edge metric, modulation complexity score (MCS), mean MLC gap, mean MLC speed, and plan modulation (PM).ResultsPTV D98% showed high robustness on average to most errors with the exception of a bank shift of 1.0 mm and large rotational errors ≥1.0° for either the couch or collimator. However, in some cases, errors close to or within generally accepted machine tolerances resulted in clinically relevant impacts. The greatest impact upon normal brain V12Gy, mean dose to normal brain, and D0.03cc brainstem was found for DLG error in alignment with other recent studies. All delivery errors had on average a minimal impact across these parameters. Comparing plans from the Monaco TPS and the Eclipse TPS, showed a lesser increase to V12Gy, mean dose to normal brain, and D0.03cc brainstem for Monaco plans (p < 0.01) when DLG error was simulated. Monaco plans also correlated to lower plan complexity. Using Spearman's correlation coefficient (r) a strong negative correlation (r ≤ -0.8) was found between the mean MLC gap and dose to OARs for DLG errors.ConclusionsReducing MLC complexity and using larger mean MLC gaps is recommended to improve plan robustness and reduce sensitivity to delivery and modelling errors. For cases in which the calculated dose distribution or dose indices are close to the clinically acceptable limits, this is especially important.
DOI del artículo:10.1002/mp.16907
Enlace a la fuente original:https://aapm.onlinelibrary.wiley.com/doi/10.1002/mp.16907
Versión del articulo depositado:info:eu-repo/semantics/publishedVersion
Acceso a la licencia de uso:https://creativecommons.org/licenses/by/3.0/es/
Departamento:Ciències Mèdiques Bàsiques
URL Documento de licencia:https://repositori.urv.cat/ca/proteccio-de-dades/
Áreas temáticas:Radiology, nuclear medicine and imaging
Radiology, nuclear medicine & medical imaging
Medicine (miscellaneous)
Medicina ii
Medicina i
Interdisciplinar
General medicine
Engenharias iv
Engenharias ii
Ciências biológicas iii
Ciências biológicas ii
Ciências biológicas i
Ciência da computação
Biotecnología
Biophysics
Astronomia / física
Antropologia / arqueologia
Palabras clave:Radiotherapy, intensity-modulated
Radiotherapy planning, computer-assisted
Radiotherapy dosage
Radiosurgery
Radiometry
Plan robustness
Plan complexity
Organs at risk
Linear accelerator-based radiotherapy
Intra-cranial stereotactic radiosurgery
Humans
Delivery errors
Brain neoplasms
Entidad:Universitat Rovira i Virgili
Fecha de alta del registro:2024-09-07