Improving the modelling of a multi-leaf collimator with tilted leaf sides used in radiotherapy

  • Identification data

    Identifier: imarina:9366473
  • Authors:

    Hussein, M
    Angerud, A
    Saez, J
    Bogaert, E
    Lemire, M
    Barry, M
    Patallo, IS
    Shipley, D
    Clark, CH
    Hernandez, V
  • Others:

    Author, as appears in the article.: Hussein, M; Angerud, A; Saez, J; Bogaert, E; Lemire, M; Barry, M; Patallo, IS; Shipley, D; Clark, CH; Hernandez, V
    Department: Ciències Mèdiques Bàsiques
    URV's Author/s: Hernandez Masgrau, Victor
    Keywords: Calculation errors Gap Mlc modelling System Tps commissioning Tps modelling
    Abstract: Background and purpose: Multi-leaf collimators (MLCs) with tilted leaf sides have a complex transmission behaviour that is not easily matched by radiotherapy treatment planning systems (TPSs). We sought to develop an MLC model that can accurately match test fields and clinically relevant plans at different centres. Materials and methods: Two new MLC models were developed and evaluated within a research version of a commercial TPS. Prototype I used adjusted-constant transmissions and Prototype II used variable transmissions at the tongue-and-groove and leaf-tip regions. Three different centres evaluated these prototypes for a tilted MLC and compared them with their initial MLC model using test fields and patient-specific quality-assurance measurements of clinically relevant plans. For the latter, gamma passing rates (GPR) at 2 %/2mm were recorded. Results: For the prototypes the same set of MLC parameters could be used at all centres, with only a slight adjustment of the offset parameter. For centres A and C, average GPR were >95 % and within 0.5 % GPR difference between the standard, and prototype models. In center B, prototypes I and II improved the agreement in clinically relevant plans, with an increase in GPR of 2.3 % +/- 0.8 % and 3.0 +/- 0.8 %, respectively. Conclusions: The prototype MLC models were either similar or superior to the initial MLC model, and simpler to configure because fewer trade-offs were required. Prototype I performed comparably to the more sophisticated Prototype II and its configuration can be easily standardized, which can be useful to reduce variability and improve safety in clinical practice.
    Thematic Areas: Oncology Radiation Radiology, nuclear medicine & medical imaging Radiology, nuclear medicine and imaging
    licence for use: https://creativecommons.org/licenses/by/3.0/es/
    Author's mail: victor.hernandez@urv.cat
    Author identifier: 0000-0003-3770-8486
    Record's date: 2024-05-23
    Papper version: info:eu-repo/semantics/publishedVersion
    Link to the original source: https://www.phiro.science/article/S2405-6316(24)00013-7/fulltext#%20
    Papper original source: Physics And Imaging In Radiation Oncology. 29 100543-
    APA: Hussein, M; Angerud, A; Saez, J; Bogaert, E; Lemire, M; Barry, M; Patallo, IS; Shipley, D; Clark, CH; Hernandez, V (2024). Improving the modelling of a multi-leaf collimator with tilted leaf sides used in radiotherapy. Physics And Imaging In Radiation Oncology, 29(), 100543-. DOI: 10.1016/j.phro.2024.100543
    Licence document URL: http://repositori.urv.cat/ca/proteccio-de-dades/
    Article's DOI: 10.1016/j.phro.2024.100543
    Entity: Universitat Rovira i Virgili
    Journal publication year: 2024
    Publication Type: Journal Publications
  • Keywords:

    Oncology,Radiation,Radiology, Nuclear Medicine & Medical Imaging,Radiology, Nuclear Medicine and Imaging
    Calculation errors
    Gap
    Mlc modelling
    System
    Tps commissioning
    Tps modelling
    Oncology
    Radiation
    Radiology, nuclear medicine & medical imaging
    Radiology, nuclear medicine and imaging
  • Documents:

  • Cerca a google

    Search to google scholar