Autor según el artículo: Fabregat-Sanjuan, A; Pàmies-Vilà, R; Rigo-Vidal, A; Pascual-Rubio, V
Departamento: Enginyeria Mecànica
Autor/es de la URV: Fabregat Sanjuan, Albert / Pascual Rubio, Vicenç
Palabras clave: Neurophysiology Evoked-potentials Evoked potentials Cranial landmarks 10/20 10 neurophysiology evoked potentials cranial landmarks 20
Resumen: The study aimed to compare the conventional method of electrode marking with a new system, EPlacement, to improve accuracy and reduce the time burden on health care professionals.Ten health care professionals marked mannequin heads and adult volunteers using both methods. Time, accuracy, and usability of each method were analyzed. Three neurophysiological diagnostic tests were performed on mannequin heads: reversal pattern visual evoked potential (three electrodes required); somatosensory evoked potentials from the upper and lower extremities (five electrodes required); and standard intraoperative neurophysiological monitoring for spine surgery (nine electrodes required). Precision scanning of the mannequins with structured light and a printed hull were used to determine the actual locations of the electrodes of the 10/20 system.The new method based on the EPlacement device represents an improvement on conventional tape measure (TM) marking and may be considered within the group of advanced methods such as navigation systems since it leads to improvements of 34% (1.7 mm) for electrode positions in the Nasion-Inion and Left tragus-Right tragus lines and 77% (12.5 mm) for electrode positions using the approximate method. It reduces the time spent per test by an average of 1 min compared to the TM method. Health care staff survey results show a positive feedback regarding usability of the new method.The study showed that the EPlacement device improves accuracy, reduces time, and is easy to use compared to the conventional method of electrode marking. The EPlacement method can facilitate the complex task of electrode marking and ultimately contribute to improved patient outcomes. It has the potential to be widely accepted and implemented in clinical practice.© 2023 The Authors. Brain and Behavior published by Wiley Periodicals LLC.
Áreas temáticas: Neurosciences Medicina ii Medicina i Interdisciplinar Farmacia Engenharias iv Ciências biológicas ii Ciências biológicas i Biotecnología Behavioral sciences Behavioral neuroscience
Acceso a la licencia de uso: https://creativecommons.org/licenses/by/3.0/es/
Direcció de correo del autor: vicenc.pascual@urv.cat a.fabregat@urv.cat
Identificador del autor: 0000-0002-2878-1369
Fecha de alta del registro: 2024-08-03
Versión del articulo depositado: info:eu-repo/semantics/publishedVersion
Enlace a la fuente original: https://onlinelibrary.wiley.com/doi/full/10.1002/brb3.3187
URL Documento de licencia: https://repositori.urv.cat/ca/proteccio-de-dades/
Referencia al articulo segun fuente origial: Brain And Behavior. e3187-e3187
Referencia de l'ítem segons les normes APA: Fabregat-Sanjuan, A; Pàmies-Vilà, R; Rigo-Vidal, A; Pascual-Rubio, V (2023). Comparison of electrode position marking procedures on the cranial surface. Brain And Behavior, (), e3187-e3187. DOI: 10.1002/brb3.3187
DOI del artículo: 10.1002/brb3.3187
Entidad: Universitat Rovira i Virgili
Año de publicación de la revista: 2023
Tipo de publicación: Journal Publications