Articles producció científicaCiències Mèdiques Bàsiques

Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality

  • Datos identificativos

    Identificador:  imarina:5118195
    Autores:  Geleijns, J; Salvado Artells, M; Veldkamp, W J H; Lopez Tortosa, M; Calzado Cantera, A
    Resumen:
    This study aimed at assessment of efficacy of selective in-plane shielding in adults by quantitative evaluation of the achieved dose reduction and image quality. Commercially available accessories for in-plane shielding of the eye lens, thyroid and breast, and an anthropomorphic phantom were used for the evaluation of absorbed dose and image quality. Organ dose and total energy imparted were assessed by means of a Monte Carlo technique taking into account tube voltage, tube current, and scanner type. Image quality was quantified as noise in soft tissue. Application of the lens shield reduced dose to the lens by 27% and to the brain by 1%. The thyroid shield reduced thyroid dose by 26%; the breast shield reduced dose to the breasts by 30% and to the lungs by 15%. Total energy imparted (unshielded/shielded) was 88/86 mJ for computed tomography (CT) brain, 64/60 mJ for CT cervical spine, and 289/260 mJ for CT chest scanning. An increase in image noise could be observed in the ranges were bismuth shielding was applied. The observed reduction of organ dose and total energy imparted could be achieved more efficiently by a reduction of tube current. The application of in-plane selective shielding is therefore discouraged. © Springer-Verlag 2006.
  • Otros:

    Referencia de l'ítem segons les normes APA: Geleijns, J; Salvado Artells, M; Veldkamp, W J H; Lopez Tortosa, M; Calzado Cantera, A (2006). Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality. EUROPEAN RADIOLOGY, 16(10), 2334-2340. DOI: 10.1007/s00330-006-0217-2
    Referencia al articulo segun fuente origial: EUROPEAN RADIOLOGY. 16 (10): 2334-2340
    DOI del artículo: 10.1007/s00330-006-0217-2
    Año de publicación de la revista: 2006-10-01
    Entidad: Universitat Rovira i Virgili
    Versión del articulo depositado: info:eu-repo/semantics/acceptedVersion
    Fecha de alta del registro: 2026-05-09
    Autor/es de la URV: LÓPEZ TORTOSA, MIGUEL / Salvadó Artells, Marçal
    Departamento: Ciències Mèdiques Bàsiques
    URL Documento de licencia: https://repositori.urv.cat/ca/proteccio-de-dades/
    Tipo de publicación: Journal Publications
    Autor según el artículo: Geleijns, J; Salvado Artells, M; Veldkamp, W J H; Lopez Tortosa, M; Calzado Cantera, A
    Grupo de investigación: Física i Cristal·lografia de Materials
    Áreas temáticas: Ciências biológicas i, Farmacia, General medicine, Medicina i, Medicine (miscellaneous), Radiology, nuclear medicine & medical imaging, Radiology, nuclear medicine and imaging
    Direcció de correo del autor: m.salvado@urv.cat, m.salvado@urv.cat
  • Palabras clave:

    Affordable and clean energy
    Efectes de la radiació
    Humans
    Monte carlo method
    Phantoms
    imaging
    Protecció radiològica
    Radiation dosage
    Radiation effects
    Radiation protection
    Tomografía computarizada (tc)
    Tomography
    x-ray computed
    X-ray computed tomography
    Medicine (Miscellaneous)
    Radiology
    Nuclear Medicine & Medical Imaging
    Nuclear Medicine and Imaging
    Ciências biológicas i
    Farmacia
    General medicine
    Medicina i
  • Documentos:

  • Cerca a google

    Search to google scholar