Computed tomography dose assessment for a 160 mm wide, 320 detector row, cone beam CT scanner

  • Datos identificativos

    Identificador:  imarina:5123201
    Autores:  Geleijns, J; Artells, M Salvado; Bruin, P Wde; Matter, R; Muramatsu, Y; McNitt-Gray, M F
    Resumen:
    Computed tomography (CT) dosimetry should be adapted to the rapid developments in CT technology. Recently a 160 mm wide, 320 detector row, cone beam CT scanner that challenges the existing Computed Tomography Dose Index (CTDI) dosimetry paradigm was introduced. The purpose of this study was to assess dosimetric characteristics of this cone beam scanner, to study the appropriateness of existing CT dose metrics and to suggest a pragmatic approach for CT dosimetry for cone beam scanners. Dose measurements with a small Farmer-type ionization chamber and with 100 mm and 300 mm long pencil ionization chambers were performed free in air to characterize the cone beam. According to the most common dose metric in CT, namely CTDI, measurements were also performed in 150 mm and 350 mm long CT head and CT body dose phantoms with 100 mm and 300 mm long pencil ionization chambers, respectively. To explore effects that cannot be measured with ionization chambers, Monte Carlo (MC) simulations of the dose distribution in 150 mm, 350 mm and 700 mm long CT head and CT body phantoms were performed. To overcome inconsistencies in the definition of CTDI100 for the 160 mm wide cone beam CT scanner, doses were also expressed as the average absorbed dose within the pencil chamber (). Measurements free in air revealed excellent correspondence between CTDI 300air and , while CTDI100air substantially underestimates CTDI300air. Results of measurements in CT dose phantoms and corresponding MC simulations at centre and peripheral positions were weighted and revealed good agreement between CTDI300w, and CTDI 600w, while CTDI100w substantially underestimates CTDI300w. provides a pragmatic metric for characterizing the dose of the 160 mm wide cone beam CT scanner. This quantity can be measured with the widely available 100 mm pencil ionization chamber within 150 mm long CT dose phantoms. CTDI300w measured in 350 mm long CT dose phantoms serves as an appropriate standard of reference for characterizing the dose of this CT scanner. A CT dose descriptor that is based on an integration length smaller than the actual beam width is preferably expressed as an (average) dose, such as for the 160 mm wide cone beam CT scanner, and not as CTDI100. © 2009 Institute of Physics and Engineering in Medicine.
  • Otros:

    Referencia de l'ítem segons les normes APA: Geleijns, J; Artells, M Salvado; Bruin, P Wde; Matter, R; Muramatsu, Y; McNitt-Gray, M F (2009). Computed tomography dose assessment for a 160 mm wide, 320 detector row, cone beam CT scanner. PHYSICS IN MEDICINE AND BIOLOGY, 54(10), 3141-3159. DOI: 10.1088/0031-9155/54/10/012
    Referencia al articulo segun fuente origial: PHYSICS IN MEDICINE AND BIOLOGY. 54 (10): 3141-3159
    DOI del artículo: 10.1088/0031-9155/54/10/012
    Año de publicación de la revista: 2009-05-21
    Entidad: Universitat Rovira i Virgili
    Versión del articulo depositado: info:eu-repo/semantics/acceptedVersion
    Fecha de alta del registro: 2026-05-09
    Autor/es de la URV: Salvadó Artells, Marçal
    Departamento: Ciències Mèdiques Bàsiques
    URL Documento de licencia: https://repositori.urv.cat/ca/proteccio-de-dades/
    Tipo de publicación: Journal Publications
    Autor según el artículo: Geleijns, J; Artells, M Salvado; Bruin, P Wde; Matter, R; Muramatsu, Y; McNitt-Gray, M F
    Grupo de investigación: Física i Cristal·lografia de Materials
    Áreas temáticas: Radiology, nuclear medicine and imaging, Radiology, nuclear medicine & medical imaging, Radiological and ultrasound technology, General medicine, Engineering, biomedical, Engenharias iv, Antropologia / arqueologia
    Direcció de correo del autor: m.salvado@urv.cat, m.salvado@urv.cat
  • Palabras clave:

    Sensitivity and specificity
    Reproducibility of results
    Relative biological effectiveness
    Radiometry
    Radiation dosage
    Humans
    Equipment failure analysis
    Equipment design
    Cone-beam computed tomography
    Body burden
    Engineering
    Biomedical
    Radiological and Ultrasound Technology
    Radiology
    Nuclear Medicine & Medical Imaging
    Nuclear Medicine and Imaging
    General medicine
    Engenharias iv
    Antropologia / arqueologia
  • Documentos:

  • Cerca a google

    Search to google scholar