    Resumen:
    NaI(Tl) and LaBr 3(Ce) detectors are frequently operated under unstable temperature conditions when used in an open environment. These temperature changes result in a peak shift and spectral distortion during measurement. Two methods are proposed to stabilise the measured spectra; they are applied using a software algorithm, without the necessity of adjusting the gain. Both methods are based on the experimental observation that the relative channel displacement due to temperature changes is approximately the same for all channels. The first method corrects the spectrum using experimental data obtained under controlled conditions in the laboratory, and thus it only depends on the detector temperature. The second method uses one known peak in the spectrum to correct all of the channels: the NORM 40K peak for the NaI(Tl) detector, the internal contaminant peak of 138La for the LaBr 3(Ce), or an external source when these two cannot be easily identified. © 2012 Elsevier Ltd. All rights reserved.
  • Otros:

    Referencia de l'ítem segons les normes APA: Casanovas, R; Morant, J J; Salvado, M (2012). Temperature peak-shift correction methods for NaI(Tl) and LaBr 3(Ce) gamma-ray spectrum stabilisation. RADIATION MEASUREMENTS, 47(8), 588-595. DOI: 10.1016/j.radmeas.2012.06.001
    Referencia al articulo segun fuente origial: RADIATION MEASUREMENTS. 47 (8): 588-595
    DOI del artículo: 10.1016/j.radmeas.2012.06.001
    Año de publicación de la revista: 2012-08-01
    Entidad: Universitat Rovira i Virgili
    Versión del articulo depositado: info:eu-repo/semantics/acceptedVersion
    Fecha de alta del registro: 2026-05-09
    Autor/es de la URV: Casanovas Alegre, Ramon / Morant Echevarne, Juan José / Salvadó Artells, Marçal / SALVADÓ USACH, MARIA TERESA
    Departamento: Ciències Mèdiques Bàsiques
    URL Documento de licencia: https://repositori.urv.cat/ca/proteccio-de-dades/
    Tipo de publicación: Journal Publications
    ISSN: 13504487
    Autor según el artículo: Casanovas, R; Morant, J J; Salvado, M
    Grupo de investigación: Física i Cristal·lografia de Materials
    Áreas temáticas: Radiation, Nuclear science & technology, Instrumentation, Engenharias ii, Astronomia / física
    Direcció de correo del autor: m.salvado@urv.cat, m.salvado@urv.cat, juanjose.morant@urv.cat, juanjose.morant@urv.cat, juanjose.morant@urv.cat, juanjose.morant@urv.cat
  • Palabras clave:

    Temperature dependence
    Spectra stabilisation
    Scintillation gamma-ray spectrometry
    Peak-shift correction
    Nai(tl)
    Labr (ce) 3
    Instrumentation
    Nuclear Science & Technology
    Radiation
    Engenharias ii
    Astronomia / física
  • Documentos:

  • Cerca a google

    Search to google scholar