Creation date in repository: 2024-05-13
Abstract: Neonatologists and neurophysiologist agree that there are no electrodes that allow for an adequate recording during long periods of time. As a result, amplitude integrated electroencephalography (aEEG) loses power to diagnose cerebral damage. In this work, a comparative analysis between commercial liquid-gel electrodes and the new aCUP-E was executed in other to evaluate their performance in a clinical trial with 15 neonates. Impedance, as well as signal features, were computed, highlighting the limitations of current electrodes in aEEG and concluding that aCUP-E electrode, which has been specifically designed for newborns, improves the quality of the recordings. Neonatòlegs i neurofisiòlegs coincideixen que no hi ha elèctrodes que permetin un registre adequat durant llargs períodes de temps. Com a resultat, l'electroencefalografia integrada d'amplitud (aEEG) perd potència per diagnosticar danys cerebrals. En aquest treball, es va realitzar una anàlisi comparativa entre elèctrodes comercials de gel líquid i el nou aCUP-E en altres per avaluar-ne el rendiment en un assaig clínic amb 15 nounats. Es va calcular la impedància, així com les característiques del senyal, destacant les limitacions dels elèctrodes actuals en aEEG i concloent que l'elèctrode aCUP-E, que ha estat dissenyat específicament per als nadons, millora la qualitat dels enregistraments.
Subject: Enginyeria Biomèdica
Language: en
Work's codirector: Pascual Rubio, Vicenç
Subject areas: Ingeniería Biomédica Biomedical Engineering Enginyeria Biomèdica
Department: Enginyeria Mecànica
Student: Rodríguez Ballabriga, Ángel
Academic year: 2022-2023
Title in different languages: Análisis de la electroencefalografía de amplitud integrada en neonatos: Evaluación comparativa del rendimiento de aCUP-E frente a electrodos de gel líquido Analysis of amplitude integrated electroencephalography in neonates: Assessing comparative performance of aCUP-E versus liquid gel electrodes Anàlisi de l'electroencefalografia d'amplitud integrada en nounats: Avaluació comparativa del rendiment d'aCUP-E davant d'elèctrodes de gel líquid
Work's public defense date: 2023-06-20
Access rights: info:eu-repo/semantics/openAccess
Keywords: aEEG, electrodes, impedancia del cuero cabelludos. aEEG, electrodes, scalp impedance. aEEG, elèctrodes, impedància del cuir cabellut.
Confidenciality: Si
Title in original language: Analysis of amplitude integrated electroencephalography in neonates: Assessing comparative performance of aCUP-E versus liquid gel electrodes
Project director: Fabregat Sanjuan, Albert
Education area(s): Enginyeria Biomèdica
Entity: Universitat Rovira i Virgili (URV)