Autor segons l'article: Herrero, P., Bäuerlein, P.S., Emke, E., Pocurull, E., de Voogt, P.
Versió de l'article dipositat: info:eu-repo/semantics/publishedVersion
Departament: Història i Història de l'Art
URL Document de llicència: https://repositori.urv.cat/ca/proteccio-de-dades/
Autor/s de la URV: P. Herrero, P.S. Bäuerlein, E. Emke, E. Pocurull, P. de Voogt
Resum: In this short communication we report on the technical implementations of coupling an asymmetric flow field-flow fractionation (AF4) instrument to a high resolution mass spectrometer (Orbitrap) using an atmospheric photoionisation interface. This will allow for the first time online identification of different fullerenes in aqueous samples after their aggregates have been fractionated in the FFF channel. Quality parameters such as limits of detection (LODs), limits of quantification (LOQs) or linear range were evaluated and they were in the range of hundreds ng/L for LODs and LOQs and the detector response was linear in the range tested (up to ~20. µg/L). The low detection and quantification limits make this technique useful for future environmental or ecotoxicology studies in which low concentration levels are expected for fullerenes and common on-line detectors such as UV or MALS do not have enough sensitivity and selectivity.
Entitat: Universitat Rovira i Virgili.
Any de publicació de la revista: 2014
Accès a la llicència d'ús: https://creativecommons.org/licenses/by/3.0/es/
ISSN: 0021-9673
Pàgina inicial: 277
Pàgina final: 282